ECA 450 Operation
Instruction 0024-9400
21
3. After HyperTerminal is running: click Transfer, click Capture Text,
then click Start to accept the default drive\directory\filename to which
all received data will be stored as an ASCII text file.
Note: By default, the captured text file will be stored on the hard drive at
C:\Program Files\Accessories\HyperTerminal\CAPTURE.TXT. If
desired, click Browse to select a different drive, directory and filename.
The next time data is captured, it will be stored under the same location
and filename.
Tip: Changing the filename extension to “CSV” (comma separated
variable) will allow the file to be directly opened in the spreadsheet
program. For example, double clicking the filename CAPTURE.CSV in
Windows Explorer will automatically start Microsoft Excel and load
the file.
4. Press analyzer’s MEMORY key to display the Memory menu screen.
5. Use ∧∨ keys to place cursor next to “Transmit Memory”; then press
ENTER to display the “Transmit Memory” screen.
«««««««MEMORY«««««««
Memory Directory
ª Transmit Memory
Clear Memory
6. Use ∧∨ keys to select one of the following transmit options; then press
ENTER to choose that option. What happens next depends on the option
chosen.
Transmit Memory
ª Last
Range
All
Last: Immediately transmits the last memory location that contains
data. Observe that the analyzer displays “Transmitting…” while also
displaying the record number being transmitted.
Range: When chosen, the range-select screen is displayed. By default
the range is from 0 to the last memory location that contains data. To
change the range, first press ENTER to move the cursor into the “From
record” data field; Then use the
∧, ∨, <, and > keys as necessary to
enter the number of the first record to be transmitted. Again press
ENTER, this time to save record number and move cursor back to left
of screen. Repeat this procedure to change the “To record” data field.
Start data transmission by moving cursor next to “Transmit” line, and
then pressing ENTER. Observe that the analyzer displays
“Transmitting…” while also displaying the record number that is
currently being transmitted.
All: Immediately transmits all memory locations that contain data.
Observe that the analyzer displays “Transmitting…” while also
displaying the record number that is currently being transmitted.