Instruction 24-9448 5-5
Calibration
known temperature that will be applied for
calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C),
and then use the and buttons to enter
an Applied value that exactly equals the set-
ting of the simulator.
Alternatively: Submerge probe tip into an ice-
water bath with a thermometer, wait several
minutes, and then use the and buttons
to enter an Applied value that exactly equals
the thermometer reading.
The calibration range is from 32 to 41 °F (0
to 5 °C). An attempt to calibrate outside this
range will cause the message “Applied Value
High" (or Low) to appear at the bottom of the
screen.
5. Wait until the Measured reading stabilizes,
and then press ENT to calibrate the TS-Zero
Measured value to that of the Applied value;
after which the message “Good Calibration” should briefly appear fol-
lowed by the CALIBRATE TS-SPAN screen.
TS-Span Procedure:
6. Set thermocouple simulator to 572 °F (300 °C), and then use the
and buttons to enter an Applied value that exactly equals the
setting of the simulator.
Alternatively: Submerge probe tip into a container of boiling water with
a thermometer, wait several minutes, and then use the and
buttons to enter an Applied value that exactly equals the thermometer
reading.
The calibration range is from 175 to 625 °F (80 to 330 °C). An attempt
to calibrate outside this range will cause the message “Applied Value
High” (or Low) to appear at the bottom of the screen.
7. Wait until the Measured reading stabilizes, and then press ENT to cali-
brate the TS-Span Measured value to that of the Applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATION LIST screen being re-displayed.
CALIBRATE TS-ZERO
Measured 34 °F
Applied 032 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3
CALIBRATE TS-SPAN
Measured 574 °F
Applied 572 °F
Press ENT
to calibrate
06/27/06 11:55:30 PM
PRINT
F1 F2 F3