NOTE: “Measured” is the current NO
2
reading, while
“Applied” is a known NO
2
level that will be applied for
calibration purposes.
3. Attach a 100 ppm NO
2
cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 80.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration stamped on the NO
2
cylinder.
NOTE: The calibration range is from 9 to 150ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-11. NO
2
Sensor Calibration
5. Adjust regulator for a flow meter indication of approximately 2  SCFH.
Wait until the measured reading stabilizes (approximately 3  minutes),
and then press ENT to calibrate the NO
2
measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibrat
ion WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear follow
ed by the CALIBRATION
LIST screen being re-
displayed. The sensor will now be
marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.