GA-40 T plus MAINTENANCE
9
The GA-40 T plusplus will not normally allow the instrument to be switched off if there are still traces of flue gas
in the system. The analyser will continue to operate until the signals from the gas cells are nearly zero. This
activity is shown by the message "VENTILATION REMOVE PROBE FROM FLUE".
The [OFF] key must be pressed twice to switch the instrument off completely. Pressing the key once will put
the analyser in the standby mode and the instrument will remain in this mode until it is activated again by
pressing "C" or switched off completely.
{{ Certain conditions, such as strong electro-magnetic fields can cause the analyser
to become "hung-up". The only solution to this problem is to switch the
instrument off. It is, however, possible that the analyser will no longer react to
the keyboard and can not be switched off.
Pressing the [OFF] key for a period of about
three seconds will always switch the analyser
off.